The protocol specifies a temperature range of 80-95 °C and temperatures at the upper end should increase extraction and de-crosslinking efficiency.
If you use cap strips to seal the BeatBox 96w PLATE and a thermoshaker for heating, please note the following: The temperature reached inside the wells may vary between different thermoshaker models. At very high temperatures the cap strip may burst open due to high vapor pressure of the lysis buffer. To avoid sample loss, perform a test run with lysis buffer to identify the highest possible temperature for your thermoshaker setup. We do not recommend using a heated lid (e.g. ThermoTop # 5308000003 Eppendorf). Use a conventional lid instead. If only a ThermoTop is available, reduce the temperature to 80-85°C.
A lower heating temperature may sometimes lead to lower protein extraction efficiency. To improve results, you can perform a second cycle of BeatBox and heating.
Alternatively, use a PCR cycler with a heated lid exerting pressure from the top and close the BeatBox 96w PLATE with the SILICONE MAT. This allows you to select a higher temperature for the de-crosslinking step.