Yes, this is possible. It is very important to spin down the BeatBox 96w PLATE at 300-500 rcf for 30-60 sec before removing the SILICONE MAT to avoid cross-contamination. If SP3 LYSE or high concentrations of SDS have been used in your own lysis buffer, a longer spin down time is recommended to decrease the foam, before transferring samples to clean tubes/plate.
The same applies to the BeatBox TUBES from the BeatBox Tissue Kit 24x. Here, we recommend a spin down at 1500 rcf for 30-60 sec.
NOTE: Please do not exceed the following max. centrifugation speeds to avoid plate/tube breakage and sample loss:
- BeatBox 96w PLATE with GYUTO BEAD in swing-bucket rotor: 500 rcf
- BeatBox 96w PLATE without GYUTO BEAD in swing-bucket rotor: 2250 rcf
- BeatBox TUBES with GYUTO BEAD: 1500 rcf
- BeatBox TUBES without GYUTO BEAD: 18000 rcf
Use a suitable plate carrier together with the swing-bucket rotor, e.g. PCR plate adapter Eppendorf #5825711009 for Eppendorf plate rotors.