PreON on-board lysis is the default setting when selecting the following sample types: Pellet, Liquid concentrated, or Liquid diluted.
PreON on-board lysis consists of the following steps:
- Load PreON workdeck including your samples according to the instructions in the menu.
- LYSE buffer is added to the sample.
- Samples are heated (83 °C, 10 min, shaking) to effectively extract, denature and reduce and alkylate proteins.
- NOTE: If you expect your samples to be viscous due to DNA released from the nuclei, nucleases can be used to reduce the viscosity of the samples, such as Benzonase. For more information and recommendations on the use of nuclease, see FAQ 1.25.
PreON off-board lysis should be selected in the PreON menu when:
- LYSE buffer (50 µL) has already been added to samples manually.
- Working with samples that require a different lysis temperature (PreON lysis temperature: 83 °C).
- Working with samples that require mechanical homogenization/lysis, such as rigid tissue samples.
Differences between on-board and off-board lysis on the PreON: