We recommend that you use the appropriate LYSE buffer provided with the PreOmics’ iST, iST-NHS or iST-BCT kits. If you want to use your own lysis buffer, please refer to the following table For all other buffers containing higher concentrations of salt, chaotropes, or detergents, we recommend using the SP3-iST Add-on kit in combination with the iST, iST-NHS, and iST-BCT kits (see FAQ 9.2. for more information on SP3-iST compatibility).
2X concentrated LYSE buffers have to be ordered in addition to the iST kits. Please contact us (info@preomics.com) for any further question on buffer compatibility or supply with reagents.
For urea-containing buffers: To reduce the risk of carbamylation of proteins, it is recommended not to heat samples above 37 °C in urea-containing buffers. Please contact us via info@preomics.com for recommendations on how to modify the protocol.
Yes, nucleases like benzonase are compatible with all PreOmics' LYSE-buffers, including iST LYSE, LYSE-BCT, LYSE-NHS, and SP3-LYSE. They can be used for various sample types, such as mammalian and bacteria cells, yeast, and tissues. Add the enzyme to the LYSE buffer used for the lysis step.
NOTE: Five units of benzonase per 100 µg protein input is sufficient for efficient DNA cleavage. For lysis of larger protein amounts, adjust the enzyme amount accordingly. Please note that benzonase is incompatible with lysis buffers containing EDTA.